A patient’s guide to pacing without wires
Leadless Pacemakers and the Medtronic Micra AVWhat Patients Need to Know
A complete pacemaker in a capsule about the size of a large vitamin, placed through a vein at the top of the thigh. How the Micra AV works, who it suits, and what the day of the procedure is like.
26 mm
long, roughly, and about 6.7 mm across. The Micra AV in one piece.
2 g
about what the whole device weighs, battery included
89%
AV synchrony at rest in the MARVEL 2 study, roughly, in complete AV block with normal sinus rhythm
Under 1 hr
for the whole procedure, typically
When a patient hears the word pacemaker, the picture in their head is usually the same one I had in medical school. A small metal box under the skin below the collarbone, with thin wires snaking down through a vein into the heart.
That is a traditional transvenous pacemaker, and it has worked extremely well for decades. But it has real downsides. The pocket can get infected. The wires can fracture. The shoulder area gets a permanent bump. For some patients, especially older adults with fragile veins or those with a history of device infection, the traditional design carries enough risk that we look for another option.
Over the last decade, that other option has become a reality. Leadless pacemakers are tiny self-contained devices that sit entirely inside the heart, with no wires and no chest pocket. The most widely used in the United States is the Medtronic Micra AV and its newer second-generation version, the Micra AV2. This article walks through how these devices work, who is a good candidate, what the procedure is like, and what the long-term outlook looks like.
The device
What a leadless pacemaker is
A leadless pacemaker is a complete pacing system packaged into a single capsule about the size of a large vitamin.
The Micra AV measures roughly 26 millimeters in length and 6.7 millimeters in diameter, weighs about two grams, and contains the battery, the pulse generator, the sensing electronics, and the pacing electrode all in one piece. There are no wires, no separate generator under the skin, and no chest incision.
The device is delivered through a small catheter inserted in the femoral vein at the top of the thigh. Under fluoroscopic guidance (live X-ray), the cardiologist navigates the catheter up the inferior vena cava, through the right atrium, across the tricuspid valve, and into the right ventricle. Small tines on the front of the device anchor it to the inner wall of the ventricular muscle. Once positioned and tested, the catheter is withdrawn, leaving the pacemaker in place.
The whole procedure is typically completed in under an hour. Most patients go home the same day or after a single overnight observation, with only a small puncture site at the groin to recover from rather than a chest incision.
In short, the capsule travels up a vein from the leg into the heart and anchors there, with no wires and no chest pocket.
Wires or no wires
Traditional and leadless, side by side
Tap between the two to see where each device sits and how it gets there.
- 1The pulse generator (battery and electronics) sits in a pocket under the skin below the collarbone. The pocket can get infected, and it leaves a permanent bump.
- 2Thin wires run through a vein into the heart. The wires can fracture.
- 3The wire tips touch the heart. A dual-chamber pacemaker has one in the atrium and one in the ventricle.
- 1A small puncture in the femoral vein at the top of the thigh. This is the only wound.
- 2A catheter carries the device up the inferior vena cava (the big vein from the lower body), through the right atrium and across the tricuspid valve.
- 3The capsule anchors to the inner wall of the right ventricle with small tines. The catheter comes out and the pacemaker stays.
- Where it sits
- A small metal box under the skin below the collarbone
- Wires
- Thin wires (leads) snaking down through a vein into the heart
- How it goes in
- Through a chest incision, with the box in a pocket under the skin
- What you see
- A permanent bump in the shoulder area
- Recovery
- Weeks rather than days, with shoulder and arm-lifting limits on the pocket side
Traditional pacemaker. A box under the skin with wires into the heart. It has worked extremely well for decades, but the pocket and the wires are where the problems come from.
Simplified drawings for teaching, not to scale. Your cardiologist and electrophysiologist choose the device and where it goes.
Keeping the beat in step
How the Micra AV senses the atrium without wires
Standard pacemakers come in single-chamber versions that pace one chamber and dual-chamber versions that pace both the atrium and the ventricle while keeping them coordinated.
The original Micra was a single-chamber device that paced only the right ventricle. That worked well for patients with permanent atrial fibrillation but was suboptimal for patients with normal atrial rhythm and AV block, because the heart loses the natural timing between atrial and ventricular contraction.
The Micra AV solved this problem in a clever way. It contains an internal three-axis accelerometer that detects the mechanical motion of the right atrium contracting.
- The atrium contractsThe right atrium (top right chamber) squeezes, the atrial kick.
- The capsule feels itFrom inside the ventricle, the accelerometer picks up that motion.
- A beat follows in stepThe device fires a paced beat in the ventricle that follows naturally from each atrial contraction.
This is called VDD pacing, and it restores AV synchrony without ever placing a wire in the atrium.
In the MARVEL 2 study, the algorithm achieved roughly 89 percent atrioventricular synchrony at rest in patients with complete AV block and normal sinus rhythm. The Micra AV2, approved by the FDA in 2023, refined this further with better synchrony at higher heart rates and longer battery life, while keeping the same size and overall design.
The leadless models in this article
| Model | Maker | What it does |
|---|---|---|
| Micra (original) | Medtronic | A single-chamber device that paced only the right ventricle |
| Micra AV | Medtronic | Paces the right ventricle in step with the atrium’s motion (VDD pacing), with no wire in the atrium |
| Micra AV2 | Medtronic | FDA approved in 2023. Better synchrony at higher heart rates and longer battery life, same size and design |
| Aveir DR | Abbott | FDA-approved, and the only dual-chamber leadless system. Two communicating leadless devices, one in the atrium and one in the ventricle. Aveir devices are designed to be retrieved at end of life. |
Who it suits
Who is a good candidate for a leadless pacemaker
Leadless pacemakers are not for everyone. They occupy a specific clinical niche, and choosing the right patient is one of the biggest parts of the decision.
The classic candidate
The classic candidate for a Micra AV or Micra AV2 has one of these.
- High-degree AV block
- That requires permanent pacing, where ventricular pacing dominates the device burden.
- A slow heart rate with atrial fibrillation
- Symptomatic bradycardia in the setting of atrial fibrillation, where atrial pacing is not needed.
- A past device infection
- Or a pocket complication, where avoiding a chest pocket reduces re-infection risk.
- Hard-to-use veins in the upper body
- Limited or hostile vascular access, including occluded subclavian veins, dialysis access, or prior vein damage.
- Frailty, advanced age, or chronic kidney disease
- Where less surgical exposure means less risk.
- A longer procedure would be too much
- Inability to lie still for or tolerate a longer transvenous procedure.
Typically not a candidate
Patients who have one of these.
- A mechanical tricuspid valve replacement
- An inferior vena cava filter
- That blocks femoral access.
- Severe femoral vein disease
- Anatomy that makes safe placement impossible
Patients who need full dual-chamber pacing with consistent atrial pacing, such as those with sinus node dysfunction in normal sinus rhythm, are usually still better served by a traditional dual-chamber pacemaker or by the newer Aveir DR dual-device leadless system from Abbott, which uses two communicating leadless devices, one in the atrium and one in the ventricle.
The case for going leadless
The benefits compared with traditional pacemakers
The advantages of going leadless are concrete and patient-facing.
There is no chest incision and no subcutaneous pocket, which means no visible bump, no pocket hematoma risk, and no shoulder restriction during recovery. Patients can usually return to most normal activities within a few days, and there are no lifting restrictions for the affected arm because no arm is involved.
Infection
The infection profile is the most striking difference. Traditional pacemakers carry a small but real risk of pocket infection or lead-related endocarditis, both of which can be dangerous and often require complete system removal.
In the Micra Post-Approval Registry and the multi-year follow-up data, there have been essentially no device-related infections requiring removal. This weighs heaviest for patients who have already had one device infection, who carry a substantially higher re-infection risk with another transvenous system.
No leads to fail
Lead-related complications also disappear because there are no leads. Lead fracture, insulation breach, and venous obstruction simply cannot happen with a leadless device.
Major complication rate at five years in the Micra registry, about
Roughly the rate seen with transvenous systems at three years
Five-year data from the Micra registry show a major complication rate of about 4.5 percent, which compares favorably with the roughly 8.5 percent rate seen with transvenous systems at three years.
Battery life
Battery life and what happens when it runs out
Battery longevity used to be one of the main concerns with leadless devices because the battery is small. Real-world data have been reassuring.
Manufacturer-projected battery life for the Micra, about
Median expected longevity in large registry analyses, depending on pacing burden and programming
The Micra has a manufacturer-projected battery life of about 12 years, and large registry analyses now estimate median expected longevity in the range of 14 to 17 years in clinical practice, depending on pacing burden and programming. The Micra AV2 was designed to extend battery life further than the original Micra AV, especially at higher pacing rates.
Micra: leave it, add another
When the battery does eventually deplete, the device is typically left in place and a new leadless pacemaker is implanted alongside it in a different region of the right ventricle.
Studies have shown that two and even three Micra devices can coexist in the right ventricle without interfering with each other or with cardiac function. For patients who started with a leadless device in their seventies or eighties, this often means the original device will outlast their need for replacement entirely.
Aveir: built to come out
The Aveir leadless system, by contrast, was designed with a dedicated retrieval system that allows the device to be physically removed at end of life.
Both approaches are being used in practice, and the choice is part of the broader conversation between patient and electrophysiologist.
Implant day
What the procedure looks like
From the patient’s perspective, a leadless pacemaker implant is similar to a cardiac catheterization.
Tap through the day below, or use Back and Next.
Step 1 of 9
You arrive fasting
You arrive fasting in the morning, and an IV is started.
The whole day as text
- Morning. You arrive fasting in the morning, and an IV is started.
- Sedation. You receive light sedation rather than general anesthesia in most cases.
- The thigh. The skin at the upper thigh is numbed, and a small puncture is made in the femoral vein.
- Into the heart. The delivery catheter is advanced under fluoroscopy (live X-ray) up the inferior vena cava, through the right atrium, across the tricuspid valve and into the right ventricle.
- Anchor and test. Small tines on the front of the device anchor it to the inner wall of the ventricle. The device is deployed and tested, and the catheter is withdrawn, leaving the pacemaker in place.
- Dressing. A pressure dressing is placed at the groin. The whole procedure is typically completed in under an hour.
- Bedrest. After a few hours of bedrest to make sure the femoral access site is sealed, most patients are walking and ready for discharge.
- Home. Many go home the same day. Most go home the same day or after a single overnight observation.
- The next few days. For a few days, avoid heavy lifting and strenuous lower-body exertion. The arms and shoulders are free.
The post-procedure restrictions are short. Patients are asked to avoid heavy lifting and strenuous lower-body exertion for a few days to protect the femoral access site, but there are no upper-body restrictions and no need to keep an arm in a sling.
What can go wrong
Risks worth discussing
No procedure is without risk. These are the risks specific to leadless pacemaker implantation.
Cardiac perforation
Uncommon, but it can cause pericardial effusion (fluid around the heart) or tamponade, and may require pericardial drainage or surgery.
Femoral access complications
Bleeding, hematoma, or vascular injury at the groin occur in a small percentage of cases.
Device dislodgement
Rare with current Micra devices because the tines anchor securely to the trabeculated (ridged) ventricular wall.
Rates of these complications have come down substantially as operator experience and device design have matured. The five-year Micra data show that the bulk of complications occur during or shortly after the implant, with very few late events. For most appropriately selected patients, the risk-benefit calculation favors leadless over transvenous, especially when infection risk or vascular access is a concern.
After the implant, when to get help
Call 911
- Signs of a stroke
- Face drooping, arm weakness, or trouble speaking.
- Chest pain
- Sudden shortness of breath
- Fainting
- Fluid can collect around the heart after a leadless implant, so fainting needs emergency care.
- Groin bleeding that won’t stop
- After 10 minutes of firm pressure.
Call the office
- A small bruise that isn’t growing
- Mild soreness at the groin
Choosing a device
How I discuss this decision in clinic
When a patient comes to me needing a permanent pacemaker, the first questions are not about the device. They are about the rhythm.
The questions I start with
- Do they need only ventricular pacing, or do they need atrial pacing too?
- Are they in permanent atrial fibrillation?
- How active are they?
- Is there any history of infection or vascular problems?
Pick a situation below to see where the article lands.
The Micra AV or Micra AV2 is often a strong first choice
In permanent atrial fibrillation, atrial pacing isn’t needed, so a single capsule in the ventricle covers the job.
- Micra AV or Micra AV2leadless, one capsule in the ventricleNamed for this
- Aveir DRleadless, two capsules, atrium and ventricle
- Traditional dual-chamber pacemakerbox under the collarbone, wires to both chambers
These pairings come from this article. The decision is made with you, the electrophysiologist who will do the implant, and often your primary care doctor.
All four situations as text
For a patient with permanent atrial fibrillation who needs rate-control pacing or for a patient with high-degree AV block who is in normal sinus rhythm, the Micra AV or Micra AV2 is often a strong first choice. For a younger active patient with isolated sinus node dysfunction, a traditional dual-chamber pacemaker or the Aveir DR may be more appropriate. For a patient with a recent device infection or hostile upper-body venous access, a leadless device often becomes the preferred option even when the rhythm picture would otherwise favor a dual-chamber system.
This is a decision I make jointly with the patient, the electrophysiologist who will perform the implant, and often with the primary care doctor who knows the patient’s broader health picture. The conversation is detailed because the device, once implanted, is in place for many years.
If you’re still sorting out whether you need a pacemaker or a defibrillator, my pacemaker or defibrillator guide covers the difference.
Common questions
Frequently asked questions
Will I be able to feel the leadless pacemaker?
No. Because the device sits inside the right ventricle and there is no pocket under the skin, there is nothing to feel from the outside. Patients cannot see or touch the device, and there is no visible bump on the chest.
Can I have an MRI with a Micra device?
Yes. Both the Micra AV and Micra AV2 are MRI-conditional, meaning MRI scans can be performed safely under specific conditions. You will need to inform the MRI technologist that you have the device so they can use the appropriate protocol.
Do leadless pacemakers have all the features of traditional pacemakers?
Most of the rate-responsive and pacing features are present, including AV synchrony in the case of the Micra AV. Single-leadless devices do not provide atrial pacing, which is a limitation for patients with sinus node dysfunction. Dual-chamber leadless systems like the Aveir DR address that gap.
How long does the battery last?
The manufacturer projects about 12 years of battery life, and registry data suggest median real-world longevity in the 14 to 17 year range depending on how often the device paces. For most older patients, the battery will not need replacement during their lifetime.
What happens when the battery does run out?
For Micra devices, the standard approach is to leave the depleted device in place and implant a new leadless pacemaker alongside it in a different part of the right ventricle. Studies have shown this is safe and effective. Aveir devices are designed to be retrieved at end of life using a dedicated snare system.
Is the procedure painful?
Most patients describe minimal discomfort. Light sedation is used during the procedure, and the only real soreness afterward is at the small femoral puncture site. There is no chest incision and no shoulder pain.
Can I exercise after a leadless pacemaker?
Yes, and usually sooner than after a traditional pacemaker. Lower-body activity should be limited for the first few days to allow the femoral site to heal, but there are no arm or shoulder restrictions and no long-term activity limitations once recovery is complete.
How do I know if I am a candidate?
The decision involves your rhythm pattern, your vascular access, your overall health, and your personal preferences. A discussion with your cardiologist and an electrophysiologist who implants both leadless and traditional pacemakers will sort out which option fits your situation best.
The bottom line
Leadless pacemakers represent one of the most meaningful technical advances in pacing in the last fifty years. The Medtronic Micra AV and Micra AV2 give us a way to provide reliable ventricular and AV-synchronized pacing in a self-contained capsule that lives entirely inside the heart, with no wires and no chest pocket. For the right patient, this means a faster, less invasive procedure, a much lower risk of infection, and a recovery that is measured in days rather than weeks. The technology is not a universal replacement for traditional pacemakers, but it is now a first-line option for many patients I see in clinic.
References
- Steinwender C, et al. “Atrioventricular Synchronous Pacing Using a Leadless Ventricular Pacemaker: Results From the MARVEL 2 Study.” JACC: Clinical Electrophysiology. 2020.
- El-Chami MF, et al. “Leadless pacemakers at 5-year follow-up: the Micra transcatheter pacing system post-approval registry.” European Heart Journal. 2024.
- Reynolds D, et al. “A Leadless Intracardiac Transcatheter Pacing System.” New England Journal of Medicine. 2016.
- Knops RE, et al. “A Dual-Chamber Leadless Pacemaker (Aveir DR i2i Study).” New England Journal of Medicine. 2023;388(25):2360-2370.
- Crossley GH, et al. “Two-year outcomes of Micra AV leadless pacemakers in the Micra AV CED study.” Heart Rhythm O2. 2024.
- Garweg C, et al. “Advancements in Leadless Pacemakers: What the Second-generation Micra AV2 Brings to Cardiac Care.” touchREVIEWS in Cardiology and Vascular Medicine. 2024.
- Medtronic Micra AV2 Leadless Pacemaker, Prescribing Information and Technical Manual.
- Vamos M, et al. “Leadless Micra pacemakers: Estimating long-term longevity. A real world data analysis.” International Journal of Cardiology. 2025.
Published on damianrasch.com. The above information was composed by Dr. Damian Rasch, drawing on individual insight and bolstered by digital research and writing assistance. The information is for educational purposes only and does not constitute medical advice.